Journal of Dentistry
Volume 36, Issue 6 , Pages 409-417, June 2008

Bone stress and interfacial sliding analysis of implant designs on an immediately loaded maxillary implant: A non-linear finite element study

  • Heng-Li Huang

      Affiliations

    • School of Dentistry, China Medical University, No. 91 Hsueh-Shih Road, Taichung 40402, Taiwan
    • Corresponding Author InformationCorresponding author. Tel.: +1 886 4 22053366x2307; fax: +1 886 4 22014043.
  • ,
  • Jui-Ting Hsu

      Affiliations

    • School of Dentistry, China Medical University, No. 91 Hsueh-Shih Road, Taichung 40402, Taiwan
  • ,
  • Lih-Jyh Fuh

      Affiliations

    • School of Dentistry, China Medical University, No. 91 Hsueh-Shih Road, Taichung 40402, Taiwan
  • ,
  • Ming-Gene Tu

      Affiliations

    • School of Dentistry, China Medical University, No. 91 Hsueh-Shih Road, Taichung 40402, Taiwan
  • ,
  • Ching-Chang Ko

      Affiliations

    • School of Dentistry, University of North Carolina at Chapel Hill, CB#7450, Chapel Hill, NC27599-7450, United States
  • ,
  • Yen-Wen Shen

      Affiliations

    • Dental Department, China Medical University Hospital, No2 Yu-Teh Road, Taichung 404, Taiwan

Received 17 July 2007; received in revised form 21 February 2008; accepted 21 February 2008.

Abstract 

Objectives

This study investigated the surrounding bone stress and the implant–bone interfacial sliding of implant designs and implant sizes of immediately loaded implant with maxillary sinus augmentation by using three-dimensional (3D) non-linear finite element (FE) analysis.

Methods

Twenty-four FE models including four implant designs (cylindrical, threaded, stepped and step-thread implants) and three implant dimensions (standard, long and wide threaded implants) with a bonded and three levels of frictional contact of implant–bone interfaces were analyzed. The maxillary model was constructed from computer tomography (CT) images of a human skull and all 3D implant models were created via the computer-aided design (CAD) software.

Results

The use of threaded implants decreased the bone stress and sliding distance obviously about 30% as compared with non-threaded (cylindrical and stepped) implants. Increasing the implant's length or diameter reduced the bone stress by 13–26%. Employing a immediately loaded implant with smooth machine surface (μ=0.3, μ represents frictional coefficient) increased the bone stress by 28–63% as compared with the osseointegrated (bonded interfaces) implants. Roughening the implant surface (μ>0.3) did not reduced the bone stress, however it did decrease the interfacial sliding between implant and bone.

Conclusions

For an immediately loaded implant placed with sinus augmentation, using threaded implant could decrease both the bone stress and implant–bone sliding distance which may improve the implant initial stability and long-term survival. Rough surface of implants shows no benefit to reduce the bone stress but they could lower the interfacial sliding. On the contrary, employing long or wide implants decrease the bone stress but they cannot diminish the interfacial sliding.

Keywords: Immediately loaded implants, Implant designs, Sinus augmentation, CT-imaging, Finite element analysis

To access this article, please choose from the options below

Login to an existing account or Register a new account.

  • Purchase this article for 31.50 USD (You must login/register to purchase this article)

    Online access for 24 hours. The PDF version can be downloaded as your permanent record.

  • Subscribe to this title

    Get unlimited online access to this article and all other articles in this title 24/7 for one year.

  • Claim access now

    For current subscribers with Society Membership or Account Number.

  • Visit SciVerse ScienceDirect to see if you have access via your institution.
 

PII: S0300-5712(08)00062-6

doi:10.1016/j.jdent.2008.02.015

Journal of Dentistry
Volume 36, Issue 6 , Pages 409-417, June 2008